  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77845",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution- Nationwide distribution to states of: FL, OK, VA, MI, TH, NM, MA, MO, WA, AND country of: Australia.",
      "recall_number": "Z-3080-2017",
      "product_description": "AirLife(TM) Reduced Heparin ARTERIAL BLOOD SAMPLER, REF 9025RHTR, 1ML SYRINE 25G 5/8\", LUER SLIP PROTECTIVE SHEATH, 3.2 IU, STERILE R,  Rx ONLY, CAREFUSION    Lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1\" needle), larger syringe (3mL), and larger amount of heparin (113IU), than what is intended for the kit.",
      "product_quantity": "5,400 units",
      "reason_for_recall": "Lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1\" needle), larger syringe (3mL), and larger amount of heparin (113IU), than what is intended for the kit.",
      "recall_initiation_date": "20170718",
      "center_classification_date": "20170828",
      "termination_date": "20190424",
      "report_date": "20170906",
      "code_info": "Lot Number 0001047501"
    }
  ]
}