  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-09",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99561",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-3071-2026",
      "product_description": "Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators    IV START KIT LAB DRAW-SH\t  \tCDS860014Q    IV START KIT\t  \tDYKM1000B    IV START KIT LAB DRAW-SH\t  \tCDS860014P    ULTRASOUND GUIDED PIV START KI\t  \tDYNDV2684",
      "product_quantity": "222,653 total",
      "reason_for_recall": "Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.",
      "recall_initiation_date": "20260720",
      "center_classification_date": "20260831",
      "report_date": "20260909",
      "code_info": "CDS860014Q    UDI-DI 10198459629822    Lots 26ABR359   26BBP789   26CBL940   26DBK987   26FBK255   26FBK816       DYKM1000B    UDI-DI 10889942508143    Lots 24KMD782   24LMH404   25AMD765   25BMD744    CDS860014P    UDI-DI 10888277797819    Lots 24IBA155   24JBY045   24LBM014   25HBR509   25JBX951    DYNDV2684    UDI-DI 10198459185465    Lot 24LBP243",
      "more_code_info": ""
    }
  ]
}