{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Genova",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86282",
      "recalling_firm": "ESAOTE S.P.A",
      "address_1": "Via Enrico Melen 77",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-3053-2020",
      "product_description": "AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems:  MyLabAlpha\t(101740000)  MyLabGamma\t(101741000)  MyLabOmega\t(101740004)  MyLabSigma\t(101741004)  MyLabSigma elite\t(101741007)  MyLabAlphaVET\t(101740001)  MyLabGammaVET\t(101741001)  MyLabDelta\t(100743000)  MyLabDeltaVET\t(100743001)  MyLabOmegaVET\t(101740005)  MyLabSigmaVET\t(101741005)",
      "product_quantity": "989 total adapters in the US",
      "reason_for_recall": "The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting.    This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.",
      "recall_initiation_date": "20200908",
      "center_classification_date": "20200930",
      "termination_date": "20230417",
      "report_date": "20201007",
      "code_info": "Serial Numbers: Below K17350001"
    }
  ]
}