{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fort Worth",
      "address_1": "6201 South Fwy",
      "reason_for_recall": "Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.",
      "address_2": "",
      "product_quantity": "367 units",
      "code_info": "Serial Numbers:  C0002,C2006,C2020,C2026,C2054,C3009R,C3010R,C3024,C3036,C3066,C3067,C3085R,C3090,C3093,C3108,C3136,C3153,C4001,C4003,C4005,C4006,C4007,C4009,C4010,C4011,C4012,C4013,C4014,C4015,C4016,C4017,C4019,C4020,C4021,C4022,C4023,C4024,C4025,C4026,C4027,C4028,C4029,C4030,C4031,C4035,C4036,C4037,C4038,C4039,C4040,C4043,C4045,C4046R,C4047R,C4049R,C4050R,C4052R,C4053R,C4054R,C4055R,C4056R,C4057R,C4058R,C4059R,C4060R,C4061R,C4062R,C4065R,C4066R,C4067R,C4068R,C4069R,C4070R,C4071R,C4072R,C4073R,C4074R,C4075R,C4076R,C4077R,C4078R,C4079R,C4080R,C4083R,C4085R,C4086R,C4087RC4088R,C4089R,C4090R,C4091R,C4092R,C4093R,C4094R,C4095R,C4096R,C4097R,C4098R,C4099R,C4100R,C4101R,C4102R,C4103R,C4104R,C4105R,C4106R,C4107R,C4108R,C4109R,C4110R,C4111R,C4112R,C4113R,C4115R,C4117R,C4118R,C4119R,C4120R,C4121R,C4122R,C4123R,C4124R,C4126R,C4127R,C4128R,C4129R,C4130R,C4132R,C4133R,C4134R,C4135R,C4136R,C4138R,C4140,C4142,C4143,C4144,C4145,C4146,C4147,C4148,C4149,C4150,C4151,C4152R,C4153,C4154,C4155,C4156,C4157R,C4158,C4159,C4160R,C4161R,C4162R,C4163R,C4164R,C4165R,C4166R,C4167R,C4168R,C4169R,C4170R,C4173R,C4174R,C4175,C4177R,C4178R,C4179R,C4180R,C4181R,C4182R,C4183R,C4184R,C4185R,C4186R,C4187R,C4188R,C4189R,C4190R,C4191R,C4192R,C4193R,C4195R,C4196R,C4197R,C4198,C4199R,C4200R,C4201,C4202R,C4203R,C4204R,C4205R,C4206R,C4207R,C4208R,C4209R,C4210R,C4211R,C4212R,C4213R,C4214R,C4215R,C4216R,C4217R,C4218R,C4219R,C4220R,C4221R,C4222R,C4223R,C4224R,C4225R,C4226R,C4227R,C4228R,C4229R,C4230R,C4231R,C4232R,C4233R,C4234R,C4235R,C4238R,C4239R,C4240R,C4241R,C4242R,C4243R,C4244R,C4245R,C4246R,C4247R,C4248R,C4249R,C4251R,C4252R,C4253R,C4254R,C4255R,C4256R,C4257R,C4258R,C4259R,C4260R,C4261R,C4262R,C4264R,C4266R,C4267R,C4268R,C4269R,C4271R,C4272R,C4273R,C4274R,C4275R,C4276R,C4278R,C4279R,C4280R,C4283R,C4284R,C4285R,C4286R,C4287R,C4288R,C4289R,C4290R,C4291R,C4292R,C4293R,C4294R,C4295R,C4296R,C4297R,C4298R,C4299R,C4300R,C4301R,C4302R,C4303R,C4304R,C4305R,C4306R,C4307R,C4308R,C4309R,C4310R,C4311R,C4312R,C4315R,C4316R,C4317R,C4318R,C4319R,C4320R,C4321R,C4323R,C4324R,C4326R,C4327R,C4328R,C4331R,C4332R,C4334R,C4337R,C4339R,C4340R,C4341R,C4342R,C4343R,C4344R,C4345R,C4346R,C4347R,C4348R,C4349R,C4352R,C4354R,C4355R,C4356R,C4357R,C4358R,C4359R,C4360R,C4361R,C4363R,C4364R,C4365R,C4366R,C4367R,C4368R,C4371R,C4372R,C4373R,C4374R,C4376R,C4377R,C4378R,C4379R,C4380R,C4381R,C4382R,C4383R,C4384R,C4385R,C4387R,C4388R,C4389R,C4394,C4396,C4397,C4399,C4406R,C4407R,C4408R,C4415,C4418,C4420,C4421,C4422,C4423,C4424,C4425,C4426",
      "center_classification_date": "20170821",
      "distribution_pattern": "Worldwide distribution.  US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom",
      "state": "TX",
      "product_description": "ORA System with VerifEye, Catalog Number 8065998300  For use during intraocular lens surgery",
      "report_date": "20170830",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Alcon Research, Ltd.",
      "recall_number": "Z-3050-2017",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "77823",
      "termination_date": "20190827",
      "more_code_info": "",
      "recall_initiation_date": "20170630",
      "postal_code": "76134-2099",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}