{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-09",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99531",
      "recalling_firm": "Diagnostica Stago, Inc.",
      "address_1": "5 Century Dr",
      "address_2": "N/A",
      "postal_code": "07054-4607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.",
      "recall_number": "Z-3047-2026",
      "product_description": "STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074;   qualitative and quantitative factor deficiency test",
      "product_quantity": "5220 units",
      "reason_for_recall": "A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified",
      "recall_initiation_date": "20260729",
      "center_classification_date": "20260828",
      "report_date": "20260909",
      "code_info": "UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027",
      "more_code_info": ""
    }
  ]
}