{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bothell",
      "address_1": "11720 N Creek Pkwy N Ste 100",
      "reason_for_recall": "Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.",
      "address_2": "",
      "product_quantity": "4 units",
      "code_info": "Reference No. 110-0021; Shipping P/N P007021-03; Device P/N P006982-03; Serial No. VAS10-0134, VAS13-0440, VAS12-0203, 000TWY.",
      "center_classification_date": "20170821",
      "distribution_pattern": "US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.",
      "state": "WA",
      "product_description": "VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA    The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).",
      "report_date": "20170830",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Solta Medical Inc",
      "recall_number": "Z-3047-2017",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "77655",
      "termination_date": "20201130",
      "more_code_info": "",
      "recall_initiation_date": "20170601",
      "postal_code": "98011-8244",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}