  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86335",
      "recalling_firm": "Ameditech Inc",
      "address_1": "9940 Mesa Rim Rd",
      "address_2": "N/A",
      "postal_code": "92121-2910",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US - Nationwide  OUS - Australia, France, Germany, New Zealand, Singapore and Spain",
      "recall_number": "Z-3036-2020",
      "product_description": "ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description):    IMCA-5AB-W / \"ImmuTest Drugs of Abuse Cup   AMP1000+BZO300+COC300+OPI2000+THC50   With Cr, Ni, pH, Bl, S.G.\",    IMC-8BBCE/ \"ImmuTest Drugs of Abuse Cup   8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300  NO ADULTERATION\",    IMCA-8CM5 /\"ImmuTest Drugs of Abuse Cup   (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50  with Cr, Ni, pH, Bl, S.G.\"",
      "product_quantity": "27,375 units",
      "reason_for_recall": "Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.",
      "recall_initiation_date": "20200818",
      "center_classification_date": "20200930",
      "termination_date": "20240930",
      "report_date": "20201007",
      "code_info": "ALL Lots Within Expiry"
    }
  ]
}