  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86335",
      "recalling_firm": "Ameditech Inc",
      "address_1": "9940 Mesa Rim Rd",
      "address_2": "N/A",
      "postal_code": "92121-2910",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US - Nationwide  OUS - Australia, France, Germany, New Zealand, Singapore and Spain",
      "recall_number": "Z-3034-2020",
      "product_description": "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description):    MF25/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10\", QF09/\"Fastect¿ II Drug Screen Dipstick Test, 4 Drug Detection: THC50+COC300+OPI300+BZO300\",    MF20/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MTD300\",  MF21/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: TCA1000,  MF23/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100\",  MF23/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100\",  MF18/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300,  MF18/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300\",    QF12/ \"Fastect¿ II Drug Screen Dipstick Test 4 Drug Detection: THC50+COC300+OPI300+MET500\",    MF10/ \"Fastect¿ II Drug Screen Dipstick Test THC50\",   MF25/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10\",  MF16 /\"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MET500\",  MF23 /\"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100\",  MF14/ \"Fastect¿ II Drug Screen Dipstick Test AMP1000\",  MF11/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: COC300\",  MF23 / Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100\",  MF25/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10\",  MF25/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10\",  MF23/ \"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100\"",
      "product_quantity": "108,000 units",
      "reason_for_recall": "Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.",
      "recall_initiation_date": "20200818",
      "center_classification_date": "20200930",
      "termination_date": "20240930",
      "report_date": "20201007",
      "code_info": "ALL Lots Within Expiry"
    }
  ]
}