  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80807",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "N/A",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide",
      "recall_number": "Z-3013-2018",
      "product_description": "CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713     Immunohistochemistry (IHC) for in vitro diagnostic use.",
      "product_quantity": "240 units",
      "reason_for_recall": "Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers.  A false negative IHC result may occur.",
      "recall_initiation_date": "20180802",
      "center_classification_date": "20180914",
      "termination_date": "20211122",
      "report_date": "20180926",
      "code_info": "Lot Numbers:  Y16507, Y23040, UDI: 4015630984398"
    }
  ]
}