{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86386",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, WA, WI and Puerto Rico.    The countries of Australia, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Czech Republic, Germany, Hungary, India, Ireland, Italy, Jordan, Lebanon, Malaysia, Mexico, Nepal, Pakistan, Panama, Portugal, Russia, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.",
      "recall_number": "Z-3006-2020",
      "product_description": "BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.",
      "product_quantity": "21455 assay systems",
      "reason_for_recall": "Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.",
      "recall_initiation_date": "20200826",
      "center_classification_date": "20200924",
      "termination_date": "20240524",
      "report_date": "20200930",
      "code_info": "Catalog Number:  33570  UDI Code:  (01)15099590231187  Serial/Lot Numbers:  921133  921348  921742"
    }
  ]
}