{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80807",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "N/A",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide",
      "recall_number": "Z-3006-2018",
      "product_description": "CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-4296      Immunohistochemistry (IHC) for in vitro diagnostic use.",
      "product_quantity": "1655 units",
      "reason_for_recall": "Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers.  A false negative IHC result may occur.",
      "recall_initiation_date": "20180802",
      "center_classification_date": "20180914",
      "termination_date": "20211122",
      "report_date": "20180926",
      "code_info": "Lot Numbers:  Y12992, Y18852, Y23051, UDI: 4015630972203"
    }
  ]
}