      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86185",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-3005-2020",
      "product_description": "Software: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK  Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180",
      "product_quantity": "64 systems US",
      "reason_for_recall": "Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location",
      "recall_initiation_date": "20200819",
      "center_classification_date": "20200923",
      "termination_date": "20230419",
      "report_date": "20200930",
      "code_info": "Serial Numbers:   100418  100419  100420  100421  100641  100642  100643  100644  100645  100704  100713  100735  100811  100812  100813  100814  101724  102013  102031  102108  102305  102369  130125  130140  130331  130335  130628  130751  130848  130874  131012  131169  131247  131284  131325  131336  131360  131503  131578  131589  131801  131966  132011  132061  132102  132108  132136  132231  132595  132729  222224  222411  222417  222475  222491  222539  222563  222601  222638  222686  222946  223004  223084  451001"
    }
  ]
}