  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86290",
      "recalling_firm": "ArthroCare Corporation",
      "address_1": "7000 W William Cannon Dr Bldg 1",
      "address_2": "N/A",
      "postal_code": "78735-8509",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV.  International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA",
      "recall_number": "Z-2999-2020",
      "product_description": "Speed Stitch Needle Cassette",
      "product_quantity": "6,366 needles",
      "reason_for_recall": "Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.",
      "recall_initiation_date": "20200817",
      "center_classification_date": "20200921",
      "report_date": "20200930",
      "code_info": "Model: OM-8850  Lot Numbers:  2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526"
    }
  ]
}