{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Milwaukee",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80928",
      "recalling_firm": "GE Medical Systems Ultrasound & Primary Care Diagnostics, LL",
      "address_1": "9900 W Innovation Dr",
      "address_2": "N/A",
      "postal_code": "53226-4856",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Puerto Rico  FOREIGN: Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Brazil, Brazil, Brunei Darussalam,  Bulgaria, Burkina Faso, Canada, Chile,   China,  Colombia, Congo, The Democratic Republic of the Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Guam, Guatemala,   Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon,  Luxembourg, Macedonia, The Former Yugoslav Republic of, Malaysia,  Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay,  Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation,  San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom,   Uruguay, and Viet Nam",
      "recall_number": "Z-2988-2018",
      "product_description": "GE Vivid E95 ultrasound system    Product The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate), Transesophageal; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, &#38; vascular).",
      "product_quantity": "12,426 (4,717 US; 7,709 OUS ALL PRODUCTS)",
      "reason_for_recall": "The system side of the power cord may break and expose the electrical conductors leading to the risk of shock.",
      "recall_initiation_date": "20180322",
      "center_classification_date": "20180906",
      "termination_date": "20210824",
      "report_date": "20180912",
      "code_info": "Each consignee have one serial numbers assigned"
    }
  ]
}