{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78411",
      "recalling_firm": "Medline Industries, Inc.",
      "address_1": "1170 S Northpoint Blvd",
      "address_2": "N/A",
      "postal_code": "60085-6757",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "CA, NM",
      "recall_number": "Z-2980-2018",
      "product_description": "LASER CDS",
      "product_quantity": "144 kits",
      "reason_for_recall": "Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as \"ChloraPrep\" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled \"ChloraPrep\". ChloraPrep\" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.",
      "recall_initiation_date": "20170412",
      "center_classification_date": "20180906",
      "termination_date": "20190805",
      "report_date": "20180912",
      "code_info": "Model: CDS980970D; Lots: 16FD2090, 16GD2340, 16ID0038, 16KD1940, 17BD1440, 17CD2208"
    }
  ]
}