{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78411",
      "recalling_firm": "Medline Industries, Inc.",
      "address_1": "1170 S Northpoint Blvd",
      "address_2": "N/A",
      "postal_code": "60085-6757",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "CA, NM",
      "recall_number": "Z-2977-2018",
      "product_description": "PICC Tray, STERILE, 12 packs each per case.",
      "product_quantity": "144 kits",
      "reason_for_recall": "Multiple surgical packs were assembled with a non-sterile CHG disk labeled as \u001cBIOPATCH\" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.",
      "recall_initiation_date": "20170412",
      "center_classification_date": "20180906",
      "termination_date": "20190805",
      "report_date": "20180912",
      "code_info": "Item number: DYNJ49177C.  Lots: 16HB4700, 16UB5076,   17MB6837"
    }
  ]
}