{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Freiburg Im Breisgau",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80628",
      "recalling_firm": "Pro-Med Instruments Gmbh",
      "address_1": "Botzinger Str. 38",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan",
      "recall_number": "Z-2971-2018",
      "product_description": "DORO LUCENT Locking Transitional Member, REF 1101.031  Component of headrest system used in open and percutaneous craniotomies and spinal surgeries",
      "product_quantity": "31 units",
      "reason_for_recall": "The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.",
      "recall_initiation_date": "20180705",
      "center_classification_date": "20180917",
      "termination_date": "20201119",
      "report_date": "20180926",
      "code_info": "Serial Numbers:  1704002  1704004  1704005  1704007  1704011  1704041  1704010  1704012  1704019  1704037  1704014  1704016  1704017  1704021  1704024  1704025  1704036  1704038  1704018  1704008  1704034  1704026  1704027  1704028  1704029  1704030  1704032  1704035  1704006  1704037  1704039  1709014"
    }
  ]
}