  "results": [
    {
      "status": "Terminated",
      "city": "Grove City",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86218",
      "recalling_firm": "Tosoh Bioscience Inc",
      "address_1": "3600 Gantz Rd",
      "address_2": "N/A",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.",
      "recall_number": "Z-2968-2020",
      "product_description": "Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy  Part number: 020661    The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay",
      "product_quantity": "39 boxes",
      "reason_for_recall": "Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.",
      "recall_initiation_date": "20200807",
      "center_classification_date": "20200914",
      "termination_date": "20210621",
      "report_date": "20200923",
      "code_info": "Lot Numbers: J803507, JX03508, JZ03509, A203510, A403511"
    }
  ]
}