  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86338",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution.",
      "recall_number": "Z-2959-2020",
      "product_description": "Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.",
      "product_quantity": "100",
      "reason_for_recall": "Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector.  This intermittent connection may result in false air bubble alarms.",
      "recall_initiation_date": "20200825",
      "center_classification_date": "20200911",
      "termination_date": "20210129",
      "report_date": "20200923",
      "code_info": "UDI: 00886799000038;    Serial Numbers: 20288  20289  20290  20291  20292  20293  20294  20295  20296  20299  20300  20301  20302  20306  20307  20308  20309  20310  20312  20317  20318  20320  20321  20322  20323  20324  20326  20328  20329  20331  20332  20334  20338  20339  20341  20342  20343  20344  20345  20346  20347  20348  20349  20350  20351  20352  20358  20365  20368  20373  20374  20375  20381  20382  20383  20384  20385  20387  20388  20392  20393  20395  20396  20397  20398  20399  20401  20402  20403  20404  20405  20408  20410  20411  20412  20555  20556  20557  20558  20559  20560  20561  20562  20563  20564  20565  20566  20567  20568  20569  20570  20571  20572  20573  20574  20575  20576  20577  20578  20579"
    }
  ]
}