      "status": "Terminated",
      "city": "Wuxi",
      "state": "N/A",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80693",
      "recalling_firm": "GE Medical Systems China Co., Ltd.",
      "address_1": "National Hi-Tech",
      "address_2": "No. 19 Changjiang Road; Dev. Zone",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-2954-2018",
      "product_description": "GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems",
      "product_quantity": "310 units",
      "reason_for_recall": "On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position.  On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.",
      "recall_initiation_date": "20180724",
      "center_classification_date": "20180830",
      "termination_date": "20201216",
      "report_date": "20180905",
      "code_info": "(a) 620, A1:   Serial Numbers:  SM618060027MA  SM618160055MA  SM618160049MA  SM618120029MA  SM618120027MA  SM618120026MA  SM618120028MA  SM618190086MA  SM618190098MA  SM618190095MA  SM618190094MA  SM618190093MA  SM618190096MA  SM618190097MA  SM618060028MA  SM618060025MA  SM618060024MA  SM618060023MA  SM618060019MA  SM618060015MA  SM618060016MA  SM618060017MA  SM618130008WA  SM618130010WA  SM618130009WA  SM618130007WA  SM618130011WA  SM618130012WA  SM618200003WA  SM618200002WA  SM618200004WA  SM618200005WA  SM618120030MA  SM618120031MA  SM618120036MA  SM618160011MA  SM618160010MA  SM618120032MA  SM618120034MA  SM618120033MA  SM618120035MA  SM618120037MA  SM618100002WA  SM618100003WA  SM618100004WA  SM618100001WA  SM618100005WA  SM618100006WA  SM618100007WA  SM618100008WA  SM618100009WA  SM618100010WA  SM618100011WA  SM618100012WA  SM618160071MA  SM618210084WA  SM618070008WA  SM618070009WA  SM618070010WA  SM618070007WA  SM618070011WA  SM618070006WA  SM618200028WA  SM618200027WA  SM618150006WA  SM618150007WA  SM618150005WA  SM618150003WA  SM618150004WA  SM618160092WA  SM618080005WA  SM618080003WA  SM618080004WA  SM618050010WA  SM618050009WA  SM618080006WA  SM618210028WA  SM618210026WA  SM618210027WA  SM618150018WA  SM618150019WA  SM618150008WA  SM618150009WA  SM618150010WA  SM618150011WA  SM618150012WA  SM618150043WA  SM618150044WA  SM618150045WA  SM618150042WA  SM618150046WA  SM618230034WA  SM618230035WA  SM618240002WA  SM618230033WA  SM618230036WA  SM618230037WA  SM618240003WA  SM618240004WA  SM618240005WA  SM618240006WA  SM618210020WA  SM618210021WA  SM618210018WA  SM618210019WA  SM618210022WA  SM618210005WA  SM618210006WA  SM618210007WA  SM618210004WA  SM618210008WA  SM618160094WA  SM618160095WA  SM618160093WA  SM618160096WA  SM618160097WA  SM618160098WA  SM618160099WA  SM618160100WA  SM618160101WA  SM618150024WA  SM618150025WA  SM618150026WA  SM618150023WA  SM618150027WA  SM618150028WA  SM618150029WA  SM618150030WA  SM618150031WA  SM618150032WA  SM618070005WA  SM618060002WA  SM618150015WA  SM618150016WA  SM618150013WA  SM618150014WA  SM618150017WA  SM618150020WA  SM618150021WA  SM618150022WA  SM618060047MA  SM618060049MA  SM618060051MA  SM618060043MA  SM618060044MA  SM618060046MA  SM618060048MA  SM618060050MA  SM618060045MA  SM618060052MA  SM618190076MA  SM618190078MA  SM618190077MA  SM618190082MA  SM618190075MA  SM618190079MA  SM618190080MA  SM618190081MA  SM618160039MA  SM618160043MA  SM618160045MA  SM618160040MA  SM618160044MA  SM618160042MA  SM618160048MA  SM618160046MA  SM618160047MA  SM618160041MA  SM618160036MA  SM618160035MA  SM618160034MA  SM618210056MA  SM618210057MA  SM618210064MA  SM618190073MA  SM618210067MA  SM618190063MA  SM618190068MA  SM618190064MA  SM618210070MA  SM618210065MA  SM618190069MA  SM618190067MA  SM618190065MA  SM618210069MA  SM618190066MA  SM618210068MA  SM618190070MA  SM618210066MA  SM618190072MA  SM618190071MA  SM618210071MA  SM618190074MA  SM618210032MA  SM618210048MA  SM618210030MA  SM618210054MA  SM618210055MA  SM618210033MA  SM618210029MA  SM618210046MA  SM618210042MA  SM618210036MA  SM618210039MA  SM618210052MA  SM618210031MA  SM618210053MA  SM618210049MA  SM618210035MA  SM618210038MA  SM618210037MA  SM618210043MA  SM618210045MA  SM618210047MA  SM618210040MA  SM618210034MA  SM618210041MA  SM618210051MA  SM618210044MA  SM618210050MA  SM618230018WA  SM618240013WA  SM618170020WA  SM618140010WA  SM618190028WA  SM618110012WA  SM618160012MA  SM618060006WA  SM618230006WA  SM618230016WA  SM618230017WA  SM618200014WA  SM618200012WA  SM618200013WA  SM618200015WA  SM618200020WA  SM618200019WA  SM618110004WA  SM618080019WA  SM618200017WA  SM618200016WA  SM618140006WA  SM618140004WA  SM618140005WA  SM618110003WA  SM618110005WA  SM618110006WA  SM618110001WA  SM618110002WA  SM618230005WA  SM618200018WA  SM618080018WA  SM618060076WA  SM618060008WA  SM618060073WA  SM618060074WA  SM618060075WA  SM618060077WA  SM618060078WA  SM618060064MA  SM618060063MA  SM618080001WA  SM618150041WA  SM618190026WA  SM618200006WA  SM618160104WA  SM618110013WA  SM618220061WA  SM618060026MA  SM618060072WA  SM618090008WA  SM618110007WA  SM618060079WA  SM618060080WA  SM618090009WA  SM618150001WA  SM618250004WA  SM618220005WA  SM618220006WA  SM618210001WA  SM618230010WA  SM618190031WA  SM618110008WA  SM618230011WA  SM617220006WA  SM618080007WA  SM618080015WA  SM618220004WA  SM618190112WA  SM618210009WA  SM618200024WA  SM618240009WA  SM618070001WA  SM618220013WA  SM618090006WA  SM618090007WA  SM618090010WA  SM618060071WA  SM618210003WA  SM618240010WA  SM618230014WA  SM618060066MA  SM618060065MA  SM618060067MA  SM618060007WA  SM618190027WA  SM618080011WA  SM618080012WA  SM618080013WA  SM618080014WA  SM618190057MA  SM618060030MA  SM618060031MA  SM618060034MA  SM618060036MA  SM618060039MA  SM618060037MA  SM618060041MA  SM618060038MA  SM618060040MA  SM618060035MA  SM618060042MA  SM618060032MA  SM618060033MA  SM618060055MA  SM618060058MA  SM618060057MA  SM618060059MA  SM618060056MA  SM618150036WA  SM618150033WA  SM618150034WA  SM618150035WA  SM618150037WA  SM618150038WA  SM618150039WA  SM618150040WA  SM618180007WA  SM618180006WA  SM618180005WA  SM618120019MA  SM618120020MA  SM618120021MA  SM618120018MA  SM618160020MA  SM618160021MA  SM618190038MA  SM618190035MA  SM618190034MA  SM618190037MA  SM618190036MA  SM618190040MA  SM618190039MA  SM618080017WA  SM618160103WA  SM618160102WA  SM618080016WA  SM618060053MA  SM618060054MA  SM618160038MA  SM618090011WA  SM618160107WA  SM618200023WA  SM618210088WA  SM618210085WA  SM618210086WA  SM618210087WA  SM618210025WA  SM618210023WA  SM618210024WA  SM618210012WA  SM618210011WA  SM618110009WA  SM618190021WA  SM618130001WA  SM618250002WA  SM618250001WA  SM618190111WA  SM618080008WA  SM618210010WA  SM618110010WA  SM618080010WA  SM618230012WA  SM618230013WA  SM618230001WA  SM618230002WA  SM618230007WA  SM618190029WA  SM618150002WA  SM618140009WA  SM618140007WA  SM618140008WA  SM618140002WA  SM618140001WA  SM618140003WA  SM618220003WA  SM618210092WA  SM618210090WA  SM618210091WA  SM618230015WA  SM618090012WA  SM618040004WA  SM618040003WA  SM618190025WA  SM618190022WA  SM618190023WA  SM618190024WA  SM618140011WA  SM618190018WA  SM618070004WA  SM618060003WA  SM618070003WA  SM618220012WA  SM618080009WA  SM618090005WA  SM618090003WA  SM618090004WA  SM618240008WA  SM618240007WA  SM618220062WA  SM618160108WA  SM618060029MA  SM618060062MA  SM618220014WA  SM618220015WA  SM618220016WA  SM618220017WA  SM618220018WA  SM618220019WA  SM618220020WA  SM618190113WA  SM618190114WA  SM618150047WA  SM618070014WA  SM618070012WA  SM618070013WA  SM618190115WA  SM618180028WA  SM618180029WA  SM618180030WA  SM618180027WA  SM618190001WA  SM618190002WA  SM618190003WA  SM618190007WA  SM618190004WA  SM618190005WA  SM618190006WA  SM618190008WA  SM618190009WA  SM618190010WA  SM618190011WA  SM618190012WA  SM618190013WA  SM618190014WA  SM618190015WA  SM618190016WA  SM618160116WA  SM618160110WA  SM618160111WA  SM618160115WA  SM618170007WA  SM618160117WA  SM618160118WA  SM618170006WA  SM618170008WA  SM618170009WA  SM618170013WA  SM618170010WA  SM618170011WA  SM618170012WA  SM618170014WA  SM618170015WA  SM618170016WA  SM618170017WA  SM618170018WA  SM618170019WA  SM618220066WA  SM618220065WA  SM618220063WA  SM618220064WA  SM618220067WA  SM618220002WA  SM618220001WA  SM618200009WA  SM618200007WA  SM618200008WA  SM618180009WA  SM618180008WA  SM618230004WA  SM618230003WA  SM618200011WA  SM618200010WA  SM618230020WA  SM618230021WA  SM618230023WA  SM618230019WA  SM618230022WA  SM618230024WA  SM618230025WA  SM618230026WA  SM618230027WA  SM618230028WA  SM618230029WA  SM618230030WA  SM618120025MA  SM618120024MA  SM618120023MA  SM618120022MA  SM618160017MA  SM618160059MA  SM618160058MA  SM618160068MA  SM618160067MA  SM618180020WA  SM618180024WA  SM618180014WA  SM618180015WA  SM618180016WA  SM618180021WA  SM618180022WA  SM618180023WA  SM618180025WA  SM618180026WA  SM618180001WA  SM618170048WA  SM618170081WA  SM618170086WA  SM618170091WA  SM618170092WA  SM618180002WA  SM618180003WA  SM618180004WA  SM618180017WA  SM618180018WA  SM618180019WA  SM618170083WA  SM618170078WA  SM618170079WA  SM618170082WA  SM618170084WA  SM618170085WA  SM618170087WA  SM618170088WA  SM618170089WA  SM618170090WA  SM618170071WA  SM618170056WA  SM618170069WA  SM618170070WA  SM618170073WA  SM618170074WA  SM618170075WA  SM618170076WA  SM618170077WA  SM618170080WA  SM618170051WA  SM618170047WA  SM618170049WA  SM618170050WA  SM618170052WA  SM618170057WA  SM618170053WA  SM618170054WA  SM618170055WA  SM618170058WA  SM618170062WA  SM618170059WA  SM618170060WA  SM618170061WA  SM618170063WA  SM618170064WA  SM618170065WA  SM618170066WA  SM618170067WA  SM618170068WA  SM618170046WA  SM618170043WA  SM618170044WA  SM618170039WA  SM618170037WA  SM618170038WA  SM618170027WA  SM618170028WA  SM618170029WA  SM618170030WA  SM618170031WA  SM618170040WA  SM618170041WA  SM618170042WA  SM618170045WA  SM618170022WA  SM618160109WA  SM618160119WA  SM618170021WA  SM618170023WA  SM618170024WA  SM618170025WA  SM618170026WA  SM618200025WA  SM618200026WA  SM618160025MA  SM618190033MA  SM618190032MA  SM618190041MA  SM618190044MA  SM618190043MA  SM618190042MA  SM618210058MA  SM618210061MA  SM618210060MA  SM618210063MA  SM618210062MA  SM618210059MA  SM618200034MA  SM618200029MA  SM618200031MA  SM618200030MA  SM618200032MA  SM618200037MA  SM618200033MA  SM618200035MA  SM618200036MA    (b) 620, A2  Serial Numbers:  SM618110017WA  SM618110018WA  SM618110015WA  SM618110016WA  SM618110019WA  SM618160004WA  SM618160106WA  SM618160105WA  SM618160001WA  SM618160002WA  SM618160003WA  SM618160005WA  SM618160006WA  SM618160007WA  SM618160008WA  SM618160074WA  SM618160009WA  SM618160072WA  SM618160073WA  SM618160075WA  SM618160076WA  SM618160077WA  SM618160078WA  SM618160079WA  SM618160080WA  SM618160081WA  SM618160082WA  SM618160083WA  SM618160084WA  SM618160085WA  SM618160086WA  SM618160087WA  SM618160088WA  SM618160089WA  SM618250044WA  SM618250043WA    (c) 650, A1  Serial Numbers:  SM718220050MA  SM718220049MA  SM718220047MA  SM718220056MA  SM718130036MA  SM718130035MA  SM718130034MA  SM718130037MA  SM718130033MA  SM718130038MA  SM718160041MA  SM718220057MA  SM718130032MA  SM718130031MA  SM718160040MA  SM718240011MA  SM718240014MA  SM718240009MA  SM718240013MA  SM718240016MA  SM718240010MA  SM718240012MA  SM718240015MA  SM718240017MA  SM718240018MA  SM718160067MA  SM718160066MA  SM718160069MA  SM718160068MA  SM718120012MA  SM718120013MA  SM718250011MA  SM718250010MA  SM718250009MA  SM718160024MA  SM718250001MA  SM718250012MA  SM718120017MA  SM718220046MA  SM718220055MA  SM718150004WA  SM718150003WA  SM718190014WA  SM718180029WA  SM718230020WA  SM718220015WA  SM718210010WA  SM718120003WA  SM718120002WA  SM718220017WA  SM718220016WA  SM718060023WA  SM718060022WA  SM718140008WA  SM718140009WA  SM718140010WA  SM718140007WA  SM718190033WA  SM718190034WA  SM718190030WA  SM718190031WA  SM718190032WA  SM718100003WA  SM718100002WA  SM718150018WA  SM718190015WA  SM718140001WA  SM718110042WA  SM718190007WA  SM718190006WA  SM718190002WA  SM718190001WA  SM718160057MA  SM718100018WA  SM718120010MA  SM718160058MA  SM718160059MA  SM718210009MA  SM718090035WA  SM718060016WA  SM718060015WA  SM718140013WA  SM718140012WA  SM718200004WA  SM718200007WA  SM718200001WA  SM718200002WA  SM718200003WA  SM718200005WA  SM718200006WA  SM718190038WA  SM718190037WA  SM718140016WA  SM718140017WA  SM718140014WA  SM718140015WA  SM718140018WA  SM718140019WA  SM718140020WA  SM718060024WA  SM718060025WA  SM718060026WA  SM718060027WA  SM718060028WA  SM718060013WA  SM718060014WA  SM718140011WA  SM718060100MA  SM718130041MA  SM718160039MA  SM718150015WA  SM718150014WA  SM718140021WA  SM718230005WA  SM718180032WA  SM718240040WA  SM718150002WA  SM718150001WA  SM718130083WA  SM718180033WA  SM718180030WA  SM718180026WA  SM718180027WA  SM718190021WA  SM718200009WA  SM718140033WA  SM718140032WA  SM718080006WA  SM718080009WA  SM718130060MA  SM718130059MA  SM718130058MA  SM718160055MA  SM718160052MA  SM718160056MA  SM718160054MA  SM718160053MA  SM718130087WA  SM718130086WA  SM718140024WA  SM718110030WA  SM718110028WA  SM718240008WA  SM718220068WA  SM718220069WA  SM718220070WA  SM718220071WA  SM718230001WA  SM718190020WA  SM718140023WA  SM718090037WA  SM718020004WA  SM718140025WA  SM718090036WA  SM718110029WA  SM718220066WA  SM718220067WA  SM718090038WA  SM718020005WA  SM718070005WA  SM718160004WA  SM718160002WA  SM718160003WA  SM718230011WA  SM718140006WA  SM718110010WA  SM71824002OWA  SM718240019WA  SM718240021WA  SM718240022WA  SM718140028WA  SM718160010WA  SM718220038WA  SM718220036WA  SM718220037WA  SM718220035WA  SM718220023WA  SM718220034WA  SM718090011WA  SM718090010WA  SM718090008WA  SM718090009WA  SM718090012WA  SM718090013WA  SM718090014WA  SM718150011WA  SM718150012WA  SM718150013WA  SM718150010WA  SM718140027WA  SM718160007WA  SM718060029WA  SM718230019WA  SM718230018WA  SM718080005WA  SM718220005WA  SM718220009WA  SM718220002WA  SM718220003WA  SM718220004WA  SM718220006WA  SM718220007WA  SM718220008WA  SM718220010WA  SM718060021WA  SM718110004WA  SM718110003WA  SM718090063WA  SM718060002WA  SM717220001WA  SM718090040WA  SM718110013WA  SM718150009WA  SM718160075WA  SM718210020WA  SM717230008WA  SM718080013WA  SM718230002WA  SM718210031WA  SM718230010WA  SM718230008WA  SM718230009WA  SM718210029WA  SM718210030WA  SM718210032WA  SM718110005WA  SM718190023WA  SM718220011WA  SM718220014WA  SM718130085WA  SM718220013WA  SM718220012WA  SM718060114WA  SM718230007WA  SM718230006WA  SM718070004WA  SM718060107WA  SM718060106WA  SM718090039WA  SM717230009WA  SM718110027WA  SM718060101WA  SM718060001WA  SM718070006WA  SM718210027WA  SM718140026WA  SM718130074WA  SM718190016WA  SM718080010WA  SM718150008WA  SM717220011WA  SM718160001WA  SM718210021WA  SM718090007WA  SM718230003WA  SM718080014WA  SM718230016WA  SM718220040WA  SM718220039WA  SM718190013WA  SM718120018MA  SM718130026MA  SM718130028MA  SM718130027MA  SM718130029MA  SM718160045MA  SM718160046MA  SM718160048MA  SM718160044MA  SM718160047MA  SM718210013WA  SM718060030WA  SM718230004WA  SM718160078WA  SM718110014WA  SM718070009WA  SM718130084WA  SM718230014WA  SM718230013WA  SM718220043WA  SM718220042WA  SM718220041WA  SM718220044MA  SM718220045MA  SM718200012WA  SM718160013MA  SM718160016MA  SM718160072WA  SM718060102WA  SM718110044WA  SM718110043WA  SM718110037WA  SM718140004WA  SM718140005WA  SM718140002WA  SM718140003WA  SM718220063MA  SM718160076WA  SM718160077WA  SM718200019WA  SM718090062WA  SM718200022WA  SM718200020WA  SM718200021WA  SM718090061WA  SM718110009WA  SM718110007WA  SM718110008WA  SM718200023WA  SM718130005WA  SM718210001WA  SM718190004WA  SM718190003WA  SM718090066WA  SM718230015WA  SM718210012WA  SM718110011WA  SM718210028WA  SM718210019WA  SM718080002WA  SM718080004WA  SM718070017WA  SM718080001WA  SM718080003WA  SM718190005WA  SM718210025WA  SM718210026WA  SM718210022WA  SM718210023WA  SM718210024WA  SM718110012WA  SM718100001WA  SM718060103WA  SM718250008WA  SM718250007WA  SM718130023MA  SM718230021WA  SM718160071MA  SM718160074WA  SM718130021MA  SM718130022MA  SM718200018WA  SM718200017WA  SM718180025WA  SM718240002WA  SM718220064WA  SM718220065WA  SM718180016WA  SM718150016WA  SM718150017WA  SM718080012WA  SM718110026WA  SM718060003WA  SM718220001WA  SM718060105WA  SM718060104WA  SM718230017WA  SM718190009WA  SM718240003WA  SM718240041WA  SM718240042WA  SM718240043WA  SM718190024WA  SM718060012WA  SM718060011WA  SM718100064WA  SM718100065WA  SM718100066WA  SM718100063WA  SM718100067WA  SM718100068WA  SM718100069WA  SM718100070WA  SM718100071WA  SM718100072WA  SM718100073WA  SM718100074WA  SM718100075WA  SM718110001WA  SM718110002WA  SM718110006WA  SM718200013WA  SM718140031WA  SM718140030WA  SM718240039WA  SM718250006WA  SM718250004WA  SM718250005WA  SM718240038WA  SM718130062WA  SM718130063WA  SM718120025WA  SM718120029WA  SM718130020WA  SM718130064WA  SM718130065WA  SM718130066WA  SM718130067WA  SM718130068WA  SM718130069WA  SM718130070WA  SM718130071WA  SM718130072WA  SM718130073WA  SM718120026WA  SM718120030WA  SM718120023WA  SM718120024WA  SM718130002WA  SM718130011WA  SM718130012WA  SM718130013WA  SM718130014WA  SM718130015WA  SM718130016WA  SM718130017WA  SM718130018WA  SM718130019WA  SM718130061WA  SM718100055WA  SM718100052WA  SM718100053WA  SM718100054WA  SM718100058WA  SM718120021WA  SM718120027WA  SM718120032WA  SM718130001WA  SM718130003WA  SM718130004WA  SM718130007WA  SM718130008WA  SM718130009WA  SM718130010WA  SM718100059WA  SM718100060WA  SM718100056WA  SM718100057WA  SM718100061WA  SM718100062WA  SM718120008WA  SM718120009WA  SM718120022WA  SM718120028WA  SM718120031WA  SM718130024MA  SM718130025MA  SM718120019MA  SM718130030MA  SM718210006MA  SM718210003MA  SM718210004MA  SM718210007MA  SM718210005MA  SM718210008MA  SM718100013WA  SM718100014WA  SM718100017WA  SM718100012WA  SM718100015WA  SM718100016WA  SM718100019WA  SM718160022MA  SM718160020MA  SM718160019MA  SM718160021MA  SM718160064MA  SM718160065MA  SM718160061MA  SM718160062MA  SM718160038MA  SM718160037MA  SM718170001WA  SM718100044WA  SM718110049WA  SM718110032WA  SM718110031WA  SM718060010WA  SM718110023WA  SM718110016WA  SM718110017WA  SM718110019WA  SM718110015WA  SM718110018WA  SM718110020WA  SM718110021WA  SM718110022WA  SM718110024WA  SM718110025WA  SM718110033WA  SM718190028WA  SM718190029WA  SM718190026WA  SM718190027WA  SM718060009WA  SM718060007WA  SM718060008WA  SM718060109WA  SM718060108WA  SM718130088WA  SM718190025WA  SM718240001WA  SM718240006WA  SM718240007WA  SM718240004WA  SM718240005WA  SM718190017WA  SM718060004WA  SM718060005WA  SM718060006WA  SM718070002WA  SM718070003WA  SM718180006WA  SM718180005WA  SM718180007WA  SM718180004WA  SM718180008WA  SM718180009WA  SM718180010WA  SM718180011WA  SM718180012WA  SM718180013WA  SM718180014WA  SM718180015WA  SM718180017WA  SM718180018WA  SM718180019WA  SM718180020WA  SM718180021WA  SM718180022WA  SM718180023WA  SM718180024WA  SM718180001WA  SM718180002WA  SM718180003WA  SM718060115WA  SM718060116WA  SM718060117WA  SM718140022WA  SM718060033WA  SM718060034WA  SM718060031WA  SM718060032WA  SM718190011WA  SM718190010WA  SM718190012WA  SM718100023WA  SM718100021WA  SM718080011WA  SM718100020WA  SM718100028WA  SM718100034WA  SM718100035WA  SM718100037WA  SM718100040WA  SM718100029WA  SM718100024WA  SM718100025WA  SM718100027WA  SM718100031WA  SM718100032WA  SM718100033WA  SM718100036WA  SM718100041WA  SM718100045WA  SM718100038WA  SM718100022WA  SM718100026WA  SM718100030WA  SM718100039WA  SM718100042WA  SM718100043WA  SM718100046WA  SM718100047WA  SM718100048WA  SM718100005WA  SM718100007WA  SM718100009WA  SM718100004WA  SM718100010WA  SM718100011WA  SM718090044WA  SM718090041WA  SM718090042WA  SM718090043WA  SM718090045WA  SM718090046WA  SM718090047WA  SM718090048WA  SM718090049WA  SM718090050WA  SM718090051WA  SM718100008WA  SM718090052WA  SM718090053WA  SM718100006WA  SM718090018WA  SM718090015WA  SM718090016WA  SM718090017WA  SM718090019WA  SM718090020WA  SM718090021WA  SM718090022WA  SM718090023WA  SM718090024WA  SM718090025WA  SM718090026WA  SM718090027WA  SM718090028WA  SM718090029WA  SM718090030WA  SM718090031WA  SM718090032WA  SM718090033WA  SM718090034WA  SM718110036WA  SM718110034WA  SM718110035WA  SM718070012WA  SM718070011WA  SM718070014WA  SM718070010WA  SM718070013WA  SM718070015WA  SM718070016WA  SM718060113WA  SM718060110WA  SM718060111WA  SM718060112WA  SM718060121WA  SM718060118WA  SM718060119WA  SM718060120WA  SM718070001WA  SM718160050MA  SM718160049MA  SM718160051MA    (d) 650, A2  Serial Numbers:  SM718120005WA  SM718120001WA  SM718240053WA  SM718240052WA    (e) 650c, A1  Serial Numbers:  SM818170005MA  SM818170002MA  SM818170001MA  SM818190002WA  SM818190001WA  SM818150002WA  SM818150001WA  SM818120001WA  SM818220001WA"
    }
  ]
}