{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bethesda",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86189",
      "recalling_firm": "LONGHORN VACCINES AND DIAGNOSTIC",
      "address_1": "2 Bethesda  Metro Cente",
      "address_2": "N/A",
      "postal_code": "20814",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution including in the state of MD.",
      "recall_number": "Z-2952-2020",
      "product_description": "PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, LH-1-15CC - Product Usage: intended for stabilization, transportation, and inactivation of infectious unprocessed nasal washes, nasopharyngeal, oral/throat swabs* suspected of containing Influenza A virus RNA. PS-MTM is also intended for the stabilization, transportation, and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis (MTB) DNA from human samples.",
      "product_quantity": "1,936,050",
      "reason_for_recall": "The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach.",
      "recall_initiation_date": "20200603",
      "center_classification_date": "20200911",
      "report_date": "20200923",
      "code_info": "Lot Numbers: 200421A  200421B  E200427A  E200427B  E2000505B  CN200423  CN200427"
    }
  ]
}