  "results": [
    {
      "status": "Terminated",
      "city": "Bonn",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86165",
      "recalling_firm": "SICAT GMBH & CO. KG",
      "address_1": "Brunnenallee 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution",
      "recall_number": "Z-2951-2020",
      "product_description": "SICAT IMPLANT V2.0",
      "product_quantity": "853 units",
      "reason_for_recall": "A dentist found implant positions are not correctly exported from the implant planning software SICAT IMPLANT V2.0 - for the specific export format CMG.DXD.",
      "recall_initiation_date": "20200715",
      "center_classification_date": "20200910",
      "termination_date": "20240103",
      "report_date": "20200916",
      "code_info": "V2.0",
      "more_code_info": ""
    }
  ]
}