  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80617",
      "recalling_firm": "Roche Diagnostics Corporation",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: CA, NY,TX, WA, and WI.",
      "recall_number": "Z-2944-2018",
      "product_description": "Roche cobas p 612 pre-analytical system (63x)  Model: 07563116001     The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes.  The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.",
      "product_quantity": "12 units",
      "reason_for_recall": "Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.",
      "recall_initiation_date": "20180716",
      "center_classification_date": "20180827",
      "termination_date": "20190516",
      "report_date": "20180905",
      "code_info": "All units"
    }
  ]
}