  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Harleysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80581",
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "address_1": "1499 Delp Dr",
      "address_2": "N/A",
      "postal_code": "19438-2936",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution; worldwide distribution.",
      "recall_number": "Z-2940-2018",
      "product_description": "SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C  T94M T96M T98M",
      "product_quantity": "N/A",
      "reason_for_recall": "The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.",
      "recall_initiation_date": "20180712",
      "center_classification_date": "20180825",
      "termination_date": "20200702",
      "report_date": "20180905",
      "code_info": "MCDA290       MHLQ370 MHMZ270 MHSN190 MHTH950 MHTR800 MHXL900  MJAT510  MJAZ020  MJBL200 MJCV760  MJFP340 MJGC620  MJGX680 MJKN210  MJLP090  MLDF860    MHLL060  MHND620 MHPA230 MHQZ510 MHSR590 MHXJ990  MHYS090 MJDW030 MJGK930 MJGX690 MJJF710 MLBA570  MLBK460  MLCT650      MHKJ180  MHMJ970 MHPK440 MHRN870 MJBP190  MJHJ860  MLBG870"
    }
  ]
}