  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Harleysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80581",
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "address_1": "1499 Delp Dr",
      "address_2": "N/A",
      "postal_code": "19438-2936",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution; worldwide distribution.",
      "recall_number": "Z-2938-2018",
      "product_description": "Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M",
      "product_quantity": "N/A",
      "reason_for_recall": "The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.",
      "recall_initiation_date": "20180712",
      "center_classification_date": "20180825",
      "termination_date": "20200702",
      "report_date": "20180905",
      "code_info": "MHJV260 MHMJ380 MHRC840 MHXF770 MJAZ910 MJDV230  MJLF440 MLCT530  MCAH650 MCAQ240 MCAW220 MCBC440 MCCJ920 MCCP260  MCDC910 MCDG320 MCDL590 MCDW940 MCFA540 MCFG000  MCFM820 MCFR060 MCFV060 MKAA520 MKAL360 MKAN510  MKAQ280 MKAZ530  MCAA590 MCAD760 MCAK990 MCBM860 MCBQ500 MCCA150  MCCG660 MCDD810 MCDG250 MCDJ980 MCDP910 MCDV760  MKAD560 MKAL370 MKBC340     MBZS180 MCAH050 MCAL000 MCBM870 MCBT030 MCCG580  MCCP590 MCDD570 MCDR280 MCDX060 MCFQ250 MCFS300  MCFT650 MCFX880 MKAB600 MKAM870    MBZY670 MCCT820 MCFQ150 MKBB070    MJLP070 MLDH060      MHLM190 MHRX850 MHSY930 MHYG710 MJCV440   MBYR870 MBZM650 MCAH670 MCAJ630 MCAQ250 MCAQ870  MCBA060 MCBD890 MCBX380 MCDF830 MCDL020 MCDR290  MCDX070 MCFA570 MCFD860 MCFH790 MKAF320 MKAG530  MKAK390 MKAL400  MKAM880 MKBA670    MCCT830       MHLJ070 MHRD620 MHRX860 MJKJ070    MCBC830 MCFK400"
    }
  ]
}