  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Harleysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80581",
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "address_1": "1499 Delp Dr",
      "address_2": "N/A",
      "postal_code": "19438-2936",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution; worldwide distribution.",
      "recall_number": "Z-2936-2018",
      "product_description": "Duo-Flow Catheter Full Tray; MCDLT3500  MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits",
      "product_quantity": "N/A",
      "reason_for_recall": "The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.",
      "recall_initiation_date": "20180712",
      "center_classification_date": "20180825",
      "termination_date": "20200702",
      "report_date": "20180905",
      "code_info": "MKAZ210 MCBK870 MCBS910 MCBZ450 MCCP990 MCCZ440 MCDNlOO MCDQ930 MCFD710 MCFN970 MCFS730 MCAS340 MCBA830 MCBR880 MCBY910 MCCF130 MCCX150 MCDF010 MCDG980 MCDP120 MCFD720 MCFJ060 MCFM520 MCFW300 MCFZ590 MKAJ380 MCAS230 MCBL150 MCBN500 MCBZ380 MCCV010 MCDF030 MCFX080 MCFV620  MCBA170  MCDC940  MKAX240  MCAC870  MCCQ000  MCFG380  MKAQ920  MCAH920  MCCW950"
    }
  ]
}