      "status": "Terminated",
      "city": "Davie",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80621",
      "recalling_firm": "Mako Surgical Corporation",
      "address_1": "2555 Davie Rd Ste 110",
      "address_2": "N/A",
      "postal_code": "33317-7424",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "All devices were distributed in the US in the following states: AK, AL, AR, AZ, CA, CO, CT, FL, IA, ID, KS, KY, LA, MA, MD, MI, MN, MO, NE, NC, NH, NJ, NY, NV, OH, PA, RI, SC, SD, TN. TX, UT, VA, WA, WI.",
      "recall_number": "Z-2925-2018",
      "product_description": "Stryker Mako Vizadisc Hip Procedure Tracking Kit",
      "product_quantity": "19,906 total devices",
      "reason_for_recall": "Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera.",
      "recall_initiation_date": "20180724",
      "center_classification_date": "20180824",
      "termination_date": "20221011",
      "report_date": "20180905",
      "code_info": "Catalog Number 107130. Lot Numbers 17097K, 17124K, 17129K, 17136K, 17143K, 17103H.    **EXPANSION** Lot Numbers 17007K, 17013K, 17024K, 17035K, 17054K, 16354H."
    }
  ]
}