  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80554",
      "recalling_firm": "Tecan US, Inc.",
      "address_1": "9401 Globe Center Dr Ste 140",
      "address_2": "N/A",
      "postal_code": "27560-6211",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC,  and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.",
      "recall_number": "Z-2923-2018",
      "product_description": "Cortisol Saliva Luminescence Immunoassay     The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.",
      "product_quantity": "870 (USA) 180 (OUS)",
      "reason_for_recall": "Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.",
      "recall_initiation_date": "20180112",
      "center_classification_date": "20180824",
      "termination_date": "20200602",
      "report_date": "20180905",
      "code_info": "In USA  RE62119 Lot # LCL129 exp. date: 2017-11-30,   RE62119 Lot # LCL133 exp. date: 2019-03-31, and   RE62111 Lot # LCL130 exp. date: 2018-03-31.  OUS    Lot 131, Lot 132"
    }
  ]
}