  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86179",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2921-2020",
      "product_description": "Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or  electrophysiology studies, for cardiac as well as interventional Radiology.",
      "product_quantity": "110 units US",
      "reason_for_recall": "The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment  conditions, the HemoBox may stop functioning and all vital signs are lost",
      "recall_initiation_date": "20200724",
      "center_classification_date": "20200828",
      "termination_date": "20211102",
      "report_date": "20200909",
      "code_info": "Serial Numbers:  100413  100414  100417  101016  101017  101019  101020  101022  101029  101031  101036  100263  100416  102008  101083  102138  102019  102021  102025  102029  102145  102147  100346  100349  102199  100095  100097  100103  100115  100116  100117  100120  100121  102158  102167  102168  102172  102272  102068  102069  102078  102080  102093  102221  100163  100165  100306  100309  101056  101099  101100  101101  101060  101068  101069  101070  101071  101072  100064  100422  100279  100408  100410  100411  100412  101043  102162  102285  102286  102183  100231  102060  102062  102049  102051  102057  102030  102031  101008  101093  101094  100175  100176  102226  101097  102269  102273  102276  102278  102279  102283  102193  102155  102157  102174  102182  102181  102214  102216  102219  102222  102237  102242  100391  101089  101095  102274  101018  102143  102091    9/01/20 Expanded S/N:  11007641  11007641  11007641    ***9/28/20 Expanded S/N***  103053  103040  103042  103008"
    }
  ]
}