  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Diego",
      "address_1": "10020 Pacific Mesa Blvd",
      "reason_for_recall": "The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion,  interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.",
      "address_2": "",
      "product_quantity": "313740",
      "code_info": "All serial numbers manufactured from April 07, 2017 to Present.",
      "center_classification_date": "20200904",
      "distribution_pattern": "U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU      O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA",
      "state": "CA",
      "product_description": "Alaris System PC Unit Model 8015",
      "report_date": "20200916",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "CareFusion 303, Inc.",
      "recall_number": "Z-2909-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85900",
      "more_code_info": "",
      "recall_initiation_date": "20200804",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}