  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ventura",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86102",
      "recalling_firm": "Neurovision Medical Products Inc",
      "address_1": "2443 Portola Rd",
      "address_2": "N/A",
      "postal_code": "93003-7726",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT,  FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS,  NE, NY,  OH, OR,  PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.",
      "recall_number": "Z-2903-2020",
      "product_description": "Cobra x5 1-CH EMG ET Tube, 6.0mm, REF:  LTE700S-5, Rx Only, Sterile EO UDI:  B006:TE700S52$$7050918B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.",
      "product_quantity": "N/A",
      "reason_for_recall": "The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.",
      "recall_initiation_date": "20200721",
      "center_classification_date": "20200827",
      "termination_date": "20210430",
      "report_date": "20200902",
      "code_info": "Kit Lots:   073119A  082019B  022820A  041620C  062520A"
    }
  ]
}