{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Paris 12e Arrondissement",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99345",
      "recalling_firm": "TISSIUM SA",
      "address_1": "74 Rue Du Faubourg Saint-Antoine",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.",
      "recall_number": "Z-2902-2026",
      "product_description": "Brand Name: COAPTIUM CONNECT  Product Name: COAPTIUM CONNECT  Model/Catalog Number: NCOCO1575  Software Version: N/A  Product Description: In situ polymerizing peripheral nerve repair device  Component: No",
      "product_quantity": "",
      "reason_for_recall": "Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.",
      "recall_initiation_date": "20260615",
      "center_classification_date": "20260806",
      "report_date": "20260812",
      "code_info": "Lot Code: Model No NCOCO1575; UDI-DI ¿03760279380087; Lot Number 250700007, 251000031, and 260300041",
      "more_code_info": ""
    }
  ]
}