  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited.  These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles.  Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.",
      "address_2": "",
      "product_quantity": "VM   96,324 Units; VS   110,424 Units; VSV   691 Units",
      "code_info": "UDI#'s  VM3 (863077): 00884838028791  VM4 (863063): 00884838000032, (863085): 00884838028807  VM6 (863064): 00884838000049, (863065): 00884838000056,  (863086): 00884838028814  VM8 (863066): 00884838000063, (863068): 00884838000087,  (863087): 00884838028821, (863088): 00884838028838, (863317):  00884838050464",
      "center_classification_date": "20180822",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia,  France, French Guiana, French Polynesia, Finland, Guadeloupe, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Lybia, Macao, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Zambia and Zimbabwe.",
      "state": "MA",
      "product_description": "Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063  863064  863065  863066  863068  863077  863085  863086  863317  863287  863288  863289     The SureSigns VM4, VM6, and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Standard and optional parameters include: \" ECG \" Respiration \" NBP \" S2pO \" IBP \" CO2 \" Temperature",
      "report_date": "20180829",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-2895-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80645",
      "more_code_info": "",
      "recall_initiation_date": "20180713",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}