  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99389",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿",
      "recall_number": "Z-2890-2026",
      "product_description": "Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic",
      "product_quantity": "7018 units",
      "reason_for_recall": "Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.",
      "recall_initiation_date": "20260706",
      "center_classification_date": "20260805",
      "report_date": "20260812",
      "code_info": "All Lots within expiry/UDI: 00630414611099",
      "more_code_info": ""
    }
  ]
}