{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99349",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, \tVA, WI, WV, WY.",
      "recall_number": "Z-2889-2026",
      "product_description": "REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments:   - B1018-280 MicroScan autoSCAN-4 Instrument  - B1018-283 MicroScan WalkAway-40 plus Instrument  - B1018-284 MicroScan WalkAway-96 plus Instrument  - B1018-440 DxM-1040 MicroScan WalkAway Instrument  - B1018-496 DxM-1096 MicroScan WalkAway Instrument",
      "product_quantity": "",
      "reason_for_recall": "Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications.\"",
      "recall_initiation_date": "20260626",
      "center_classification_date": "20260805",
      "report_date": "20260812",
      "code_info": "UDI: 15099590799250/ V5.03 Panel Update 11",
      "more_code_info": ""
    }
  ]
}