  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hamburg",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75160",
      "recalling_firm": "Spiegelberg Gmbh & Co. KG",
      "address_1": "Tempowerkring 4",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to PA only",
      "recall_number": "Z-2888-2016",
      "product_description": "Spiegelberg Brain Pressure Monitoring System  Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P    Product Usage:  lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.",
      "product_quantity": "22 units",
      "reason_for_recall": "In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg.    The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.",
      "recall_initiation_date": "20091229",
      "center_classification_date": "20160926",
      "termination_date": "20161005",
      "report_date": "20161005",
      "code_info": "Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393;   Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122"
    }
  ]
}