  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75016",
      "recalling_firm": "R & D Systems, Inc.",
      "address_1": "614 McKinley Pl NE",
      "address_2": "N/A",
      "postal_code": "55413-2610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: CA, CO, KY, LA, MD, MA, NJ, OH, RI, UT, VA, WA.    OUS: AUSTRALIA, China, INDIA, INDONESIA, ISRAEL, JAPAN, MEXICO, SOUTH KOREA, UNITED KINGDOM.",
      "recall_number": "Z-2887-2016",
      "product_description": "R&D Systems, Quantikine¿ IVD¿sTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor (sTfR) concentrations in human serum and plasma For in Vitro Diagnostic Use, REF DTFR1.",
      "product_quantity": "725 kits (502 US, 223 OUS)",
      "reason_for_recall": "R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine¿ IVD¿ Human sTfR Immunoassay kit were out of the range high.  Internal testing within R&D Systems confirmed a problem with Control results being out of range high.",
      "recall_initiation_date": "20160824",
      "center_classification_date": "20160926",
      "termination_date": "20170726",
      "report_date": "20161005",
      "code_info": "331636, 331637, 332060, 332090, 332924, 334184, 335062, 335740, 336694, 339307, 340362."
    }
  ]
}