  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99379",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr",
      "address_2": "",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of CA and VA. The country of Japan.",
      "recall_number": "Z-2880-2026",
      "product_description": "Syva Emit Disopyramide Assay  Enzyme Immunoassay, Lidocaine  SMN 10445370  Catalog No 6N319UL",
      "product_quantity": "8 US; 44 OUS",
      "reason_for_recall": "Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.",
      "recall_initiation_date": "20260626",
      "center_classification_date": "20260804",
      "report_date": "20260812",
      "code_info": "UDI-DI 00842768001406 Lot U1",
      "more_code_info": ""
    }
  ]
}