  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Valencia",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98974",
      "recalling_firm": "Boston Scientific Neuromodulation Corporation",
      "address_1": "25155 Rye Canyon Loop",
      "address_2": "",
      "postal_code": "91355-5004",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina",
      "recall_number": "Z-2879-2026",
      "product_description": "INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500;   INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700.  part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.",
      "product_quantity": "1079",
      "reason_for_recall": "Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.",
      "recall_initiation_date": "20260617",
      "center_classification_date": "20260805",
      "report_date": "20260812",
      "code_info": "Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638.  All unused products.",
      "more_code_info": ""
    }
  ]
}