  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "The PPK System Spacer Block Alignment Tower mating feature that mates with the spacer block was undersized, which could potentially lead to mating issues between the devices. This could cause a delay in surgery of less than 30 minutes. There is an option in the surgical technique for performing limb alignment without the alignment tower.",
      "address_2": "",
      "product_quantity": "385",
      "code_info": "63465803, 63762260, 63767434, 63799794",
      "center_classification_date": "20180818",
      "distribution_pattern": "Worldwide Distribution - Nationwide distribution and the countries of Japan and Netherlands.",
      "state": "IN",
      "product_description": "Persona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35    Product Usage:  Intended to aid in preparatory surgical procedures",
      "report_date": "20180829",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-2876-2018",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "80548",
      "termination_date": "20191009",
      "more_code_info": "",
      "recall_initiation_date": "20180622",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}