{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99484",
      "recalling_firm": "Bard Access Systems, Inc.",
      "address_1": "605 N 5600 W",
      "address_2": "",
      "postal_code": "84116-3738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN.  OUS: CA, SG",
      "recall_number": "Z-2872-2026",
      "product_description": "BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK,  BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK,  BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK,  BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK,  BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK",
      "product_quantity": "80,200",
      "reason_for_recall": "After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy",
      "recall_initiation_date": "20260730",
      "center_classification_date": "20260814",
      "report_date": "20260826",
      "code_info": "REF/UDI-DI/Lot(Expiration): D015151NK/00801741163586/BSLC9275(2026-08-03), BSLC9563(2026-08-13), BSLC9567(2026-08-25), BSLC10103(2026-09-27), BSLC10516(2026-11-01), BSLC10526(2026-11-16), BSLC10872(2026-12-09), BSLC11216(2028-07-02), BSLC11456(2028-07-23);  D015251NK/00801741163678/BSLC9467(2026-08-10), BSLC9575(2026-09-20), BSLC10105(2026-10-04), BSLC10522(2026-11-07), BSLC10868(2026-12-02), BSLC11212(2028-06-18), BSLC11454(2028-07-17), BSLC11463(2028-08-07), BSLC11884(2028-09-15), BSLC11890(2028-09-18), BSLC12082(2028-10-08), BSLC12115(2028-10-20), BSLC12123(2028-10-22);  D015351NK/00801741163593/BSLC11458(2028-07-24); 00801741163678/BSLC9277(2026-08-04), BSLC9485(2026-08-12), BSLC9569(2026-08-26), BSLC10099(2026-09-07), BSLC10514(2026-10-16), BSLC10518(2026-10-21), BSLC10524(2026-11-08), BSLC10870(2026-12-03), BSLC10874(2028-06-12), BSLC11218(2028-07-07);  D015451NK/00801741163609/BSLC9279(2026-08-06), BSLC9565(2026-08-17), BSLC9571(2026-09-07), BSLC10101(2026-09-17), BSLC10456(2026-10-14), BSLC10520(2026-10-22), BSLC10866(2026-11-21), BSLC11214(2028-06-23), BSLC11450(2028-07-09), BSLC11460(2028-07-31), BSLC11873(2028-08-13), BSLC11888(2028-09-25), BSLC12007(2028-10-02), BSLC12105(2028-10-15);  D015551NK/00801741163616/BSLC9465(2026-08-07), BSLC9573(2026-09-26), BSLC10107(2026-10-02), BSLC10528(2026-11-20), BSLC11452(2028-07-10)",
      "more_code_info": ""
    }
  ]
}