  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sparks",
      "address_1": "7 Loveton Cir",
      "reason_for_recall": "A portion of this lot was manufactured using Gentamicin instead of Vancomycin. Gentamicin is an inhibitory agent of Neisseria organisms. This  product is a selective plated medium to provide enhanced growth and recovery of N  gonorrhoeae. Inhibited growth may cause a false negative result or a delayed result leading to  incorrect or delayed treatment.",
      "address_2": "BD Diagnostic Systems",
      "product_quantity": "10,360 plates",
      "code_info": "8110781",
      "center_classification_date": "20180817",
      "distribution_pattern": "US Distribution in states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, ID, IN, KS, KY, MA, MI, MS, MT, NC, JH, NY, OK, OR, PA, SC, TN, TX, WA, and WI.",
      "state": "MD",
      "product_description": "BD BBL GC-Lect Agar; Cat. No. 297715     The device is a selective medium providing enhanced growth and recovery of Neisseria gonorrhoeae and better inhibition of contamination bacteria and fungi.",
      "report_date": "20180829",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Co.",
      "recall_number": "Z-2869-2018",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "80573",
      "termination_date": "20190709",
      "more_code_info": "",
      "recall_initiation_date": "20180621",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}