  "results": [
    {
      "status": "Terminated",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80577",
      "recalling_firm": "Philips Medical Systems Gmbh, DMC",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide; International to 66 countries",
      "recall_number": "Z-2866-2018",
      "product_description": "DigitalDiagnost, Single-Detect (Stitching Patient Support) 712062",
      "product_quantity": "",
      "reason_for_recall": "The hook does not securely hold the footplate in the vertical position.  Therefore, the footplate can fall or swing down and injury operator or patient.",
      "recall_initiation_date": "20180627",
      "center_classification_date": "20180817",
      "termination_date": "20240430",
      "report_date": "20180829",
      "code_info": "Serial Numbers:   386786/SN0702512  375366  416255/SN0902004  358322  388667/SN0802105  375367  397229/SN0802250  396838/SN0802365  386108/SN0802044  405687/SN0802469  354125  362438  406156/SN0802537  413697/SN0802634  402853/SN0902012  405889/SN0802629  383385/SN0802011  378827/SN0802021  384013/SN0702495  358017  358496  377925/SN0802060  356953  355621  357888  356956  368431  356963  406016/SN0802476  346522  375370  70700  351321  392504/SN0802182",
      "more_code_info": ""
    }
  ]
}