  "results": [
    {
      "status": "Terminated",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80577",
      "recalling_firm": "Philips Medical Systems Gmbh, DMC",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide; International to 66 countries",
      "recall_number": "Z-2863-2018",
      "product_description": "DigitalDiagnost Single Detector (Stitching Patient Support) 712020",
      "product_quantity": "",
      "reason_for_recall": "The hook does not securely hold the footplate in the vertical position.  Therefore, the footplate can fall or swing down and injury operator or patient.",
      "recall_initiation_date": "20180627",
      "center_classification_date": "20180817",
      "termination_date": "20240430",
      "report_date": "20180829",
      "code_info": "Serial Numbers:   399921/SN08000023  410038/SN08000091  402577/SN09000137  479651/SN11000391  443237/SN09000411  496881/SN12000104  489345/SN11000590  459498/SN10000231  483525/SN11000377  468302/SN10000537  485692/SN11000443  422533/SN09000072  450553/SN10000089  414087/SN08000100  472611/SN11000056  443235/SN09000450  481036/SN11000471  471070/SN10000695  471047/SN10000700  455221/SN10000251  460072/SN10000336  478109/SN11000327  475856/SN11000315  415754/SN08000063  443660/SN10000010  440700/SN10000030  440260/SN09000350  447603/SN09000468  448647/SN10000111  467487/SN11000085  467484/SN10000602  477526/SN11000311  490997/SN12000106  451267/SN10000104  445433/SN09000434  488711/SN11000615  470906/SN11000142  485620/SN12000002  448671/SN10000076  480052/SN11000286  480053/SN11000288  439920/SN09000329  412942/SN08000036  418438/SN08000110  431111/SN09000237",
      "more_code_info": ""
    }
  ]
}