  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80577",
      "recalling_firm": "Philips Medical Systems Gmbh, DMC",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide; International to 66 countries",
      "recall_number": "Z-2858-2018",
      "product_description": "DigitalDiagnost Dual Detector (Stitching Patient Support) 712022",
      "product_quantity": "",
      "reason_for_recall": "The hook does not securely hold the footplate in the vertical position.  Therefore, the footplate can fall or swing down and injury operator or patient.",
      "recall_initiation_date": "20180627",
      "center_classification_date": "20180817",
      "termination_date": "20240430",
      "report_date": "20180829",
      "code_info": "Serial Numbers:   488713/SN11000577  488714/SN11000572  444623/SN10000136  480503/SN11000473  480504/SN11000478  410040/SN08000021  422832/SN09000061  462407/SN10000363  448130/SN10000008  464981/SN10000471  455752/SN10000200  454061/SN10000216  426292/SN09000386  455546/SN10000143  420090/SN09000018  474817/SN11000080  474238/SN11000042  461799/SN10000347  455621/SN10000163  471148/SN11000122  471147/SN11000120  482904/SN11000375  482905/SN11000373  482903/SN11000379  440249/SN09000347  481021/SN11000446  487808/SN11000526  441778/SN09000452  432989/SN09000285  478307/SN12000004  476338/SN12000026  476034/SN12000027  422536/SN09000199  486638/SN11000494  466808/SN10000464  462593/SN10000439  471150/SN10000660  478233/SN11000233  473897/SN11000165  482501/SN11000527  453200/SN10000133  456093/SN10000204  461979/SN10000701  461976/SN10000704  443093/SN09000454  445110/SN09000461  16750755  477524/SN11000434  475932/SN11000341  475866/SN11000350  451542/SN10000098  442534/SN09000423  443443/SN09000420  484558/SN11000525  479928/SN11000287  466029/SN10000420  472613/SN11000162  480061/SN11000474  453085/SN10000125  467927/SN10000713  450779/SN10000062  480881/SN11000274  476835/SN11000179  442005/SN10000298  444627/SN10000142  442310/SN10000149  478181/SN11000455  481142/SN11000318  467348/SN11000207  441637/SN09000354  441638/SN09000356  464863/SN10000533  464880/SN10000714  421977/SN09000115  449365/SN10000110  464755/SN10000391  430355/SN09000214  451543/SN10000100",
      "more_code_info": ""
    }
  ]
}