      "status": "Ongoing",
      "city": "Netanya",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99307",
      "recalling_firm": "THERANICA BIOELECTRONICS LTD",
      "address_1": "Poleg Industrial Park",
      "address_2": "45, Ha-Melakha",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "The product's entry point to the US is a 3PL warehouse from which it is being shipped to DME distributors or specialty pharmacy who perform the shipment to the end-user",
      "recall_number": "Z-2855-2026",
      "product_description": "Nerivio Infinity. Model/Catalog Number: FGD000100.   Software Version: FW Version 4.1.2.0, App version 7.28.0.11",
      "product_quantity": "11,886 units",
      "reason_for_recall": "Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.",
      "recall_initiation_date": "20260323",
      "center_classification_date": "20260731",
      "report_date": "20260812",
      "code_info": "Model/Catalog Number: FGD000100. UDI: 01072900175781181125120221NI2544-0300673 Software Version: FW Version 4.1.2.0, App version 7.28.0.11  GTIN: 7290017578118.  Serial Number: NI2544-0300673.",