  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86116",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-2848-2020",
      "product_description": "Cios Alpha (VA20) - Model: 10308191  Cios Alpha (VA30) - Model: 11105200  Cios Spin (VA30) -  Model: 10308194    Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications",
      "product_quantity": "US: 282 units; OUS: 1363",
      "reason_for_recall": "Hardware error- Hot Plugging  feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause  permanent loss of imaging  functionality.",
      "recall_initiation_date": "20200714",
      "center_classification_date": "20200822",
      "termination_date": "20210723",
      "report_date": "20200902",
      "code_info": "S-N Affected systems:  10xxx:  10001; 10003; 10010   10027; 10029   10031; 10033   10040; 10042   10045; 10048   10052; 10054   10110; 10112   10119    11xxx: 11000   11004; 11006   11028; 11030   11101; 11103   11183; 11185   11205; 11207   11372; 11374   11387; 11600   11608    12xxx: 12000   12014; 12016   12024; 12026   12045; 12047   12052; 12054; 12200   12207; 12301  12493; 12495   12505; 12507   12522; 12524   12550; 12552   12571; 12573; 12575   12610; 12612   12746; 12800   12803; 12805   12811; 12813   12889; 12891   12959; 12961   12973;  12975   12999     13xxx: 13000   13085; 13087; 13088; 13092; 13103   13105; 13114; 13121  4xxxx:  416; 40000; 40001; 40003   40021; 40030; 40038; 40070  5xxxx:  50001   50053; 50055; 50075    Expanded Recall: 11143, 12490"
    }
  ]
}