  "results": [
    {
      "country": "United States",
      "city": "Vernon Hills",
      "address_1": "75 N Fairway Dr",
      "reason_for_recall": "The Oral Airway pouch within the PINK/RED Oxygen Delivery Module, 7700RAW, was incorrectly packaged with the 8 cm Berman Oral Airway. The pouch contained within the module is correctly labeled as 5 cm but actually contains an 8 cm Berman Oral Airway component. Therefore, the component within the pouch is incorrect.    The PINK/RED Oxygen Delivery Module, 7700RAW, is intended to be used within the PINK/RED Pouch of the BROSELOW pediatric safety system.    Three (3) lots (lot# 914873, lot# 914876 and lot #914878) of the PINK/RED Oxygen Delivery Module, 7700RAW, are affected by this voluntary Field Safety Corrective Action.",
      "address_2": "",
      "product_quantity": "544 (2/PACKS)",
      "code_info": "Lot numbers 914873, 914876, and 914878 of the individual modules (7700RAW), which are contained in Lot number 914867 of the 2/PACK (7700RAW2). Expiration Date JAN 2021.",
      "center_classification_date": "20160920",
      "distribution_pattern": "Illinois",
      "state": "IL",
      "product_description": "Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, PINK/RED Oxygen Delivery Module 2/PACK, Product Code 7700RAW2",
      "report_date": "20160928",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Carefusion 2200 Inc",
      "recall_number": "Z-2846-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74647",
      "termination_date": "20180302",
      "more_code_info": "",
      "recall_initiation_date": "20160629",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}