{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "2850 Frontier Dr",
      "reason_for_recall": "OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due to the devices being mislabeled.",
      "address_2": "",
      "product_quantity": "41",
      "code_info": "Product Number: 00-1050-3218 \u0013 Lot Number: NM02346",
      "center_classification_date": "20160919",
      "distribution_pattern": "Nationwide Distribution to AZ, CA, CO, DE, FL, GA, KY, MI, MN, MO, KY, RI, NY, OH, OR, RI, TN, TX, VA, WA    Foreign: None    VA/DOD: None",
      "state": "IN",
      "product_description": "3.5mm Bowed Locking Compression Femur Plates, 18 hole  Orthopediatrics PediLoc Locking Plate System comes in sets that offer the advantage of both conventional and locking plate fixation devices and instrumentation in one system. Utilizing both locking and non- locking screws, PediLoc offers a construct that resists angular collapse while simultaneously acting as an effective aid to fracture reduction.",
      "report_date": "20160928",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "OrthoPediatrics Corp",
      "recall_number": "Z-2844-2016",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "74928",
      "termination_date": "20170324",
      "more_code_info": "",
      "recall_initiation_date": "20160727",
      "postal_code": "46582-7001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}