      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "The handle can separate from the MRx housing due to breakage of mounts on the rear case.",
      "address_2": "",
      "product_quantity": "394 units",
      "code_info": "US Serial Numbers:  US00100227  US00100297  US00100326  US00100337  US00100429  US00100715  US00101184  US00101468  US00101676  US00101690  US00102018  US00203809  US00205745  US00205756  US00206142  US00206498  US00206615  US00206821  US00207095  US00207101  US00207144  US00207209  US00207268  US00207295  US00207321  US00207324  US00207345  US00208019  US00208263  US00208431  US00208519  US00210744  US00210966  US00211478  US00211485  US00212939  US00213455  US00213819  US00214103  US00214173  US00214222  US00316387  US00316389  US00316444  US00317530  US00318184  US00318460  US00321017  US00321936  US00323885  US00325280  US00326630  US00327632  US00328050  US00328154  US00328968  US00330028  US00330043  US00330093  US00331414  US00332111  US00534036  US00534522  US00535148  US00535803  US00539517  US00540460  US00541251  US00541256  US00541257  US00541366  US00541627  US00543683  US00543856  US00544956  US00545221  US00545649  US00554157  US00555242  US00571887  US00574441  US00574758    Canada:  US00100770  US00101622  US00101811  US00204864  US00204987  US00205107  US00325446  US00325734  US00539650  US00565585  US00569569    ROW:  US00101691  US00208635  US00327490  US00101209  US00206875  US00208214  US00209571  US00212610  US00212964  US00100372  US00536445  US00100883  US00207958  US00100861  US00211015  US00100171  US00100628  US00534545  US00321215  US00205732  US00208194    MRx units that have the Old Cap Plate, shipped before 2010.",
      "center_classification_date": "20160919",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, Canada, AUSTRALIA, FRANCE, GERMANY, INDIA, JAPAN, NETHERLANDS, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, and UNITED KINGDOM.",
      "state": "MA",
      "product_description": "HeartStart MRx Monitor/Defibrillator  Models: M3535A and M3536A",
      "report_date": "20160928",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-2842-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75041",
      "termination_date": "20200724",
      "more_code_info": "",
      "recall_initiation_date": "20160826",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}